Kit for glycated albumin determination

上海ballbet贝搏体育西甲医药科技有限公司
  • HR
  • Contact
  • Products

    bb平台-ballbet贝搏体育西甲-ballbet贝搏体育


    Product name: bb平台-ballbet贝搏体育西甲-ballbet贝搏体育

    Method: enzymic method

    Basic index:

     

    amount of specimen: 3 ul

    dosage of R1: 240 ul

    dosage of R2: 60 ul

    dominant wavelength: 546 nm

    sub-wave length: 700 nm

    assay method: endpoint method

    direction of reaction: forward direction, up

    reaction temperature: 37℃

    scaling mode: Linear

    incubating time of R1+ sample: 300 sec

    reaction time: 120 sec

     

    Clinical significance: It is used for determination of content of glycated albumin in serum. It is clinically used for blood glucose testing for nephritic patient, anemia patient, diabetic and woman during pregnancy.

     

    Product name

    Art.No.

    Packaging specification

    Main technical performance index

     

     

     

     

    bb平台-ballbet贝搏体育西甲-ballbet贝搏体育

    (GA-L)

    GA-1070

    R1: 1 × 40 ml R2: 1 × 10 ml             ALB: 1 × 50 ml

    GA reagent:

    ◆ blank absorbency≤0.20(546nm)

    ◆sensitivity for analysis: 285umol/L △A≥0.100

    ◆ degree of accuracy: relative deviation≤10%

    ◆ degree of precision: with-in degree of precision: CV≤5%

                     between degree of precision: relative deviation≤8%

    ◆linearity range: 10~1500umol/L, r≥0.99

    ALB reagent:

    ◆ blank absorbency≤0.80(630nm)

    ◆ sensitivity for analysis: 4.5g/dL△A≥0.200

    ◆ degree of accuracy: relative deviation≤5%

    ◆ degree of precision: with-in degree of precision: CV≤5%

                    between degree of precision: relative deviation≤8%

    ◆ linearity range: 0.003~7.0g/dL, r≥0.99

    GA-2070

    R1: 2 × 40 ml R2: 2 × 10 ml             ALB : 2 × 50 ml

    GA-1060

    R1: 1 × 80 ml R2: 1 × 20 ml              ALB : 1 × 100 ml

    GA-2080

    R1: 1 × 80 ml R2: 1 × 20 ml              ALB : 1 × 100 ml

    GA-2040

    R1: 2 × 40 ml R2: 2 × 10 ml            ALB : 2 × 50 ml

     

    Packaging specification with Art. No. of 40 refers to Toshiba 40 type biochemical analyzer used packaging;

    Packaging specification with Art. No. of 60 refers to Hitachi 7150, 7060, Shimadzu CL7200, 7300, 8000, Olympus AU800, Toshiba 120FR, Abbott 2000, 8000 biochemical analyzer used packaging;

    APackaging specification with Art. No. of 70 refers to Hitachi 170A, 7170, 7080, 7180, 7600P, Olympus AU400, 600, 640, 1000, 2700, 5400, Roche Group P800, Siemens 1650, 1800, 2400 biochemical analyzer used packaging;